Ultrasonic waves experiment system of reflexion
Objects
Demonstrating the principle of an echo sounder.
Determining the velocity of sound in air from the transit time of a sound pulse and the distance to the reflecting object.
Determining distance by measuring the transit time of the sound pulse.
Principles
Ultrasonic waves are reflected at the boundary surfaces between media with differing resistances to sound waves. An echo sounder (or sonar)device emits pulsed ultrasonic signals and measures the time in which a signal is reflected from such a boundary surface to the receiver. To simplify the configuration, the transmitter and receiver are in the same location.
The time between transmission and reception can be used to determine the distance to the reflecting object (if the velocity of sound is known), or to determine the velocity of sound over a known distance. This method is commonly used e.g. to determine water depths at sea.
In the experiment, the echo-sounder principle is used to determine the velocity of sound in air, and to determine distances.
Two ultrasonic transducers serve as the transmitter and receiver, depending on their connection.
A piezoelectric body converts electrical to mechanical energy. When the AC voltage is applied to the piezoelectric body, the transducer configured as a transmitter supplies a sufficiently high sound amplitude at a resonance frequencies (approx. 40 kHz). Conversely, sound waves generate mechanical oscillations in the transducer when configured as a receiver. The amplitude of the resulting piezoelectric AC voltage is proportional to the sonic amplitude.
 | System composition |
2 pcs 1 pc 1 pc 1 pc 2 pcs 2 pcs 1 pc 1 pc | Ultrasonic transducers 40 kHz AC amplifier Generator 40 kHz Digital storage oscilloscope Test leads Stand base, V-shape Metal scale, 1 m Reflection plate | P416000 F16-015 F16-014 DQ7202CA PTL927 P101413
|
|
|
|
|
F16-014 Generator 40 kHz
Features With continuance and spacing square wave generator for operating source, for ultrasonic transducer 40 kHz (P416000) as an emitter. Inner and external frequency counter
Technical Data Generator Frequency range: 40 kHz, can be set from 35 kHz to 50 kHz Pulse operation: pulse duration approx. 0.2 ms pulse spacing approx. 80 ms Transducer output voltage: >18 Vpp Trigger output voltage: >9 Vpp Counter Frequency range: 1KHz~150kHz Sensitivity: 100mV Max. input voltage: 20V Connection sockets: 4 mm dia. Dimensions: 19 cm × 13.5 cm × 7 cm Weight: 0.5kg
|
 |
 | F16-015 AC-amplifier
Features Sensitive amplifier with microphone input for verifying ultrasonic waves in conjunction with an ultrasonic transducer (P416000) as a receiver, and sound amplification
Technical Data Gain: 10× to 1000×, continuously adjustable Frequency range: 10 kHz (100Hz microphone input) to 50 kHz Outputs: signal, trigger and level, short-circuit proof Max. signal output: 4 Vp-p Trigger output: TTL compatible Max. DC level output: 4 V Connection sockets: 4 mm dia. Dimensions: 19 cm × 13.5 cm × 7 cm Weight: 0.5kg |
P416000 Ultrasonic transducer 40 kHz
Features Piezoelectric air ultrasonic transducer for experiments in the areas of geometric and wave-mechanical acoustics. The transducer is used as transmitter and receiver. In housing, on stand rod, with coax. connection cable.
Technical Data Resonance frequency: 40 kHz Bandwidth: approx. 6 kHz Capacitance: 2000 pF Connection: 1 m coax. cable with 4 mm sockets Housing: 48 mm × 27 mm dia. Stand rod: 20 cm × 10 mm dia. | 
|